Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Soltec TA200-839 (1 year) CAL VER /5520 DATE: 10-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 24 NUMBER OF LINES: 221 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- R N K P F V 1.003 ASK+ X 1.004 HEAD [MEM2] 1.005 RSLT =[MEM2] 1.006 DISP Set RECORD : ON 1.006 DISP Disable all other channels! 1.007 MATH S[1] = MEM2 1.008 MEMI Enter [MEM2] plug-in type (DCV, TC, or RTD). 1.009 JMPL DCV_PLUG_IN ZCMPI(MEM2, "DCV") 1.010 JMPL TC_PLUG_IN ZCMPI(MEM2, "TC") 1.011 JMPL RTD_PLUG_IN ZCMPI(MEM2, "RTD") 1.012 JMP 1.008 1.013 LABEL DCV_PLUG_IN 1.014 MATH MEM2 = S[1] 1.015 SET CHART SPEED : 60mm/min 1.015 SET INPUT TYPE : DC 1.015 SET LINE WIDTH : MIDDLE 1.015 SET FILTER : 1Hz 1.015 SET GRID : 100mm 1.015 SET OPERATE : START 1.015 SET DISPLAY : DIGITAL 1.016 DISP Connect 5520A to the UUT as follows: 1.016 DISP [32] 5520A NORMAL HI to UUT [MEM2] + INPUT 1.016 DISP [32] 5520A NORMAL LO to UUT [MEM2] - INPUT 1.017 HEAD {DC VOLTAGE} 1.018 HEAD DC VOLTAGE: {1mV Range} 1.019 SET RANGE : 1mV 1.020 MESS SLEW the 5520 for a +FULL SCALE indication. 1.020 MESS 1.021 5520 1 1mV 0.2% 2W 2.001 HEAD DC VOLTAGE: {2mV Range} 2.002 SET RANGE : 2mV 2.003 5520 2 2mV 0.2% 2W 3.001 HEAD DC VOLTAGE: {5mV Range} 3.002 SET RANGE : 5mV 3.003 5520 5 5mV 0.2% 2W 4.001 HEAD DC VOLTAGE: {10mV Range} 4.002 SET RANGE : 10mV 4.003 5520 10 10mV 0.2% 2W 5.001 HEAD DC VOLTAGE: {20mV Range} 5.002 SET RANGE : 20mV 5.003 5520 20 20mV 0.2% 2W 6.001 HEAD DC VOLTAGE: {50mV Range} 6.002 SET RANGE : 50mV 6.003 5520 50 50mV 0.2% 2W 7.001 HEAD DC VOLTAGE: {100mV Range} 7.002 SET RANGE : 100mV 7.003 5520 100 100mV 0.2% 2W 8.001 HEAD DC VOLTAGE: {200mV Range} 8.002 SET RANGE : 200mV 8.003 5520 200 200mV 0.2% 2W 9.001 HEAD DC VOLTAGE: {500mV Range} 9.002 SET RANGE : 500mV 9.003 5520 500 500mV 0.2% 2W 10.001 HEAD DC VOLTAGE: {1V Range} 10.002 SET RANGE : 1V 10.003 5520 1 1V 0.2% 2W 11.001 HEAD DC VOLTAGE: {2V Range} 11.002 SET RANGE : 2V 11.003 5520 2 2V 0.2% 2W 12.001 HEAD DC VOLTAGE: {5V Range} 12.002 SET RANGE : 5V 12.003 5520 5 5V 0.2% 2W 13.001 HEAD DC VOLTAGE: {10V Range} 13.002 SET RANGE : 10V 13.003 5520 10 10V 0.2% 2W 14.001 HEAD DC VOLTAGE: {20V Range} 14.002 SET RANGE : 20V 14.003 5520 20 20V 0.2% 2W 15.001 HEAD DC VOLTAGE: {50V Range} 15.002 SET RANGE : 50V 15.003 5520 50 50V 0.2% 2W 16.001 MESS 16.002 JMP 24.001 16.003 LABEL TC_PLUG_IN 16.004 HEAD {THERMOCOUPLE} 16.005 OPBR This procedure will test one of the following types 16.005 OPBR of thermocouples: E, J, K, R, S, T, or B. 16.005 OPBR 16.005 OPBR Do you wish to test one of these thermocouples? 16.006 JMPT 16.009 16.007 RSLT =Not tested 16.008 JMP 24.001 16.009 MEM2 Enter thermocouple type (E, J, K, R, S, T, or B): 16.010 JMPL VALID_TC ZCMPI(MEM2, "E") 16.011 JMPL VALID_TC ZCMPI(MEM2, "J") 16.012 JMPL VALID_TC ZCMPI(MEM2, "K") 16.013 JMPL VALID_TC ZCMPI(MEM2, "R") 16.014 JMPL VALID_TC ZCMPI(MEM2, "S") 16.015 JMPL VALID_TC ZCMPI(MEM2, "T") 16.016 JMPL VALID_TC ZCMPI(MEM2, "B") 16.017 JMP 16.005 16.018 LABEL VALID_TC 16.019 MATH S[2] = MEM2 16.020 MATH MEM2 = S[1] 16.021 DISP Connect the 5520A TC terminals to UUT [MEM2] using 16.021 DISP proper therocouple wire and connectors. 16.022 SET CHART SPEED : 60mm/min 16.022 SET LINE WIDTH : MIDDLE 16.022 SET GRID : 100mm 16.022 SET OPERATE : START 16.022 SET DISPLAY : DIGITAL 16.023 MESS Slew the 5520A for a +FULL SCALE indication. 16.023 MESS 16.024 MATH MEM2 = S[2] 16.025 JMPL THERMOCOUPLE_TYPE_E ZCMPI(MEM2, "E") 16.026 JMPL THERMOCOUPLE_TYPE_J ZCMPI(MEM2, "J") 16.027 JMPL THERMOCOUPLE_TYPE_K ZCMPI(MEM2, "K") 16.028 JMPL THERMOCOUPLE_TYPE_R ZCMPI(MEM2, "R") 16.029 JMPL THERMOCOUPLE_TYPE_S ZCMPI(MEM2, "S") 16.030 JMPL THERMOCOUPLE_TYPE_B ZCMPI(MEM2, "B") 16.031 LABEL THERMOCOUPLE_TYPE_T 16.032 HEAD {T-type THERMOCOUPLE} 16.033 SET INPUT TYPE : T 16.033 SET TEMPERATURE LOWER : 0 degC 16.033 SET TEMPERATURE UPPER : 400 degC 16.034 5520 400.0degC 0.2% 0.2U _T TC 2W 17.001 JMP 24.001 17.002 LABEL THERMOCOUPLE_TYPE_R 17.003 HEAD {R-type THERMOCOUPLE} 17.004 SET INPUT TYPE : R 17.004 SET TEMPERATURE LOWER : 0 degC 17.004 SET TEMPERATURE UPPER : 1700 degC 17.005 5520 1700degC 0.2% 0.2U _R TC 2W 18.001 JMP 24.001 18.002 LABEL THERMOCOUPLE_TYPE_S 18.003 HEAD {S-type THERMOCOUPLE} 18.004 SET INPUT TYPE : S 18.004 SET TEMPERATURE LOWER : 0 degC 18.004 SET TEMPERATURE UPPER : 1700 degC 18.005 5520 1700degC 0.2% 0.2U _S TC 2W 19.001 JMP 24.001 19.002 LABEL THERMOCOUPLE_TYPE_B 19.003 HEAD {B-type THERMOCOUPLE} 19.004 SET INPUT TYPE : B 19.004 SET TEMPERATURE LOWER : 100 degC 19.004 SET TEMPERATURE UPPER : 1800 degC 19.005 5520 1800degC 0.2% 0.2U _B TC 2W 20.001 JMP 24.001 20.002 LABEL THERMOCOUPLE_TYPE_K 20.003 HEAD {K-type THERMOCOUPLE} 20.004 SET INPUT TYPE : K 20.004 SET TEMPERATURE LOWER : 0 degC 20.004 SET TEMPERATURE UPPER : 1300 degC 20.005 5520 1300degC 0.2% 0.2U _K TC 2W 21.001 JMP 24.001 21.002 LABEL THERMOCOUPLE_TYPE_J 21.003 HEAD {J-type THERMOCOUPLE} 21.004 SET INPUT TYPE : J 21.004 SET TEMPERATURE LOWER : 0 degC 21.004 SET TEMPERATURE UPPER : 1200 degC 21.005 5520 1200degC 0.2% 0.2U _J TC 2W 22.001 JMP 24.001 22.002 LABEL THERMOCOUPLE_TYPE_E 22.003 HEAD {E-type THERMOCOUPLE} 22.004 SET INPUT TYPE : E 22.004 SET TEMPERATURE LOWER : 0 degC 22.004 SET TEMPERATURE UPPER : 1000 degC 22.005 5520 1000degC 0.2% 0.2U _E TC 2W 23.001 JMP 24.001 23.002 LABEL RTD_PLUG_IN 23.003 HEAD {RTD} 23.004 MATH MEM2 = S[1] 23.005 SET CHART SPEED : 60mm/min 23.005 SET LINE WIDTH : MIDDLE 23.005 SET GRID : 100mm 23.005 SET OPERATE : START 23.005 SET DISPLAY : DIGITAL 23.005 SET INPUT TYPE : P 23.005 SET TEMPERATURE LOWER : 0 degC 23.005 SET TEMPERATURE UPPER : 600 degC 23.006 DISP Connect the 5520A to the UUT as follows: 23.006 DISP [32] 5520A NORMAL HI to UUT [MEM2] + INPUT 23.006 DISP [32] 5520A NORMAL LO to UUT [MEM2] - INPUT 23.007 5520 600degC 0.3% 0.2U R1 2W 24.001 SET OPERATE : STOP 24.002 END